SOP for Clean Bench
Manufacturer: Klenzaids Bioclean Devices Ltd.
A21, MIDC Street 3, Andheri (E), Mumbai 400093
Ph: 91-22-28218921, Email:
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
-
Switch on the power supply. Switch on the HEPA filter, tube-light, UV light by pressing the appropriate switch placed at the side panel of the clean bench.
-
Clean the clean bench after its use. Close the front panel if you are not working there.
-
Do not leave anything that you bring in the clean bench after your work. If you need to do so, put a note there telling about your name, the date when you kept it and when will you remove it.
-
Do not leave any solution in the beakers after your work get finished. Dispose them at the appropriate place and keep the beakers on the rack near the wash basin.
-
Switch off everything in the clean bench after you finish your work. It includes ultrasonic cleaner and variac of spin drier.
Ultrasonic Cleaner
Manufacturer: Transonic Digital
Contact: Siskin Instruments
5/5 11 Cross MSREC road, MSRIT, PO Mathikela, Extn: Bangalore
Ph: 3342908/3369560
-
Switch on the cleaner.
-
Set the time for required cleaning from the knob labelled as TIMER.
-
Set the temperature from the knob labeled as TEMPERATURE. Wait till the temperature gets stabilized
-
Start cleaning by pressing the START button. It will be operative as per the set time. If you need to finish in between, press the FINISH button.
-
Power setting should be 100. Do not go beyond 120.
Spin Drier
-
Put the sample in the groove of the spin drier.
-
Switch on the variac. Red light on the variac gets on.
-
Set the voltage on the variac. Do not go beyond 100V as it causes shaking of the clean bench.
Contact:
Sunita Mehta(9005660465)
Deepa Singh (9648049212)
|