Spin coater, DUCOM, PR-6-M2 |
SOP For Spin-Coater 1. Turn on the main switch. 2. Switch on the rotary pump. 3. Turn on the switch of spin coater display. 4. Place the substrate on the chuck. Note: The substrate size should be larger than the O-ring diameter. For smaller substrate, ask from the super user for smaller diameter O-ring chuck. 5. Set the required program · Press Select process on display. · Press edit mode · Set the number of steps required. · Set the time, rpm and acceleration (using tabs and arrows buttons) · Press run mode
6. Open the nitrogen supply valve and the set the pressure 65-70 psi. Turn the vacuum on by pressing vacuum button on display. 7. Dispense the solution/PR/Polymer on the substrate and press start. Precautions: · While dispensing or cleaning ,Chemical should not go into the motor. · While cleaning, first take the o-ring out from the chuck and then clean. · Clean the instrument properly after use.
Please maintain the log-book. Contact Details Asit Prakash Bajpai (9005018516) Sunita Mehta (9005660465)
|