Heating/ Colling Holders (HRTEM only)
For heating/ cooling stage user is requested to provide following details:
- Required Temperature
- Ramping rate of the temperature for e.g. 10°C/min
- Melting point of the sample (in degrees celcius)
- TG/DT analysis of sample
Double Tilt Heating Holder (Gatan, Model 652)
- Number of samples (maximum 2 per slot)
- Note:
- Maximum temperature 900°C
- EDS not possible (please do not request for EDS during heating of sample as it can damage the EDS detector)
- Maximum +/- 10° tilt is available
- Additional Rs.500 per hour will be charged after 3 hours if slot is extended
Cooling Stage (Gatan, Model 613)
- Only one sample can be done in one slot
- Maximum +/- 20°C can be achieved with cryo holder